← Back to CVE List

CVE-2017-12725

Published: 2018-02-15T10:29Z
Last Modified: 2024-11-21T03:10Z
Source: MITRE CVE List
License: MITRE-CVE-TOS
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection. > MITRE Terms of Use apply – see LICENSE‑MITRE.txt