← Back to CVE List

CVE-2018-8870

Published: 2018-07-03T01:29Z
Last Modified: 2024-11-21T04:14Z
Source: MITRE CVE List
License: MITRE-CVE-TOS
Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system. > MITRE Terms of Use apply – see LICENSE‑MITRE.txt