← Back to CVE List

CVE-2018-19010

Published: 2019-01-28T22:29Z
Last Modified: 2024-11-21T03:57Z
Source: MITRE CVE List
License: MITRE-CVE-TOS
Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity. > MITRE Terms of Use apply – see LICENSE‑MITRE.txt