← Back to CVE List

CVE-2019-5024

Published: 2019-04-11T18:29Z
Last Modified: 2024-11-21T04:44Z
Source: MITRE CVE List
License: MITRE-CVE-TOS
A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability. > MITRE Terms of Use apply – see LICENSE‑MITRE.txt