← Back to CVE List

CVE-2020-27252

Published: 2020-12-14T20:15Z
Last Modified: 2024-11-21T05:20Z
Source: MITRE CVE List
License: MITRE-CVE-TOS
Medtronic MyCareLink Smart 25000 all versions are vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device. > MITRE Terms of Use apply – see LICENSE‑MITRE.txt